4.11 Electric Stretcher Patient Control Switch
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 34
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
4. Peel the patient control switch (B) off the patient control unit (A) by
peeling from one of the edges.
Replacement
SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury and equipment damage.
1. Clean any adhesive left on the patient control unit (A).
2. Remove the adhesive backing off of the new patient control switch.
NOTE:
Position the patient control switch so the picture of the head is at the head end
of the stretcher.
3. Insert the wiring of the new patient control switch (B) through the patient
control unit (A), and press the switch into position.
4. Connect the patient control switch (B) to the connector.
5. After the switch installation is complete, place the cover housing (C) into
position, and fasten it with the two screws (D).
6. Plug the stretcher into an appropriate power source, and check for proper
operation.