4.1 PC-300–PC-550 Procedural/Fixed Height/Trauma/Surgical/ OB/GYN Stretcher Siderail
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 8
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
Figure 4-2. OB/GYN Stretcher Siderail
CAUTION:
Ensure that the siderail is supported on both ends. Damage to the
siderail bracket can occur if one end drops, and the other end is
positioned in the siderail bracket.
3. Support the siderail, and remove the swing pivot bolts (A).
4. Remove the siderail from the swing arms (D).
5. Remove the four Oilite®
1
bushings (B) from the swing arms (D).
Replacement
1. Install the two Oilite® bushings (B) on each swing arm (D).
2. Align the siderail swing arm bracket with the swing arm (D).
3. Insert the swing pivot bolts (A) through the siderail bracket and swing
arm (D).