2.18 Trauma Stretcher Cassette Lift Handle Malfunction
Chapter 2: Troubleshooting Procedures
Page 2 - 58
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
2.18 Trauma Stretcher Cassette Lift Handle Malfunction
1. The cassette lift cable and spring are properly attached.
Yes
No
↓
→
Make sure the round vinyl cap is attached to the open end of the
spring. Orient the spring so the open end that attaches to the
cable link is facing the base of the stretcher.
2. The cassette lift mechanism is obstructed.
Yes
No
↓
→
Go to step 4.
3. Remove the obstruction from the cassette lift mechanism or the upper
frame cassette support surface. If this solves the problem, go to “Final
Actions” on page 2-32. Otherwise, call Hill-Rom Technical Support at
(800) 445-3720.