4.23 Procedural/Electric/Trauma/Surgical/OB/GYN Stretcher Upper Frame
Chapter 4: Removal, Replacement, and Adjustment Procedures
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
Page 4 - 75
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
4
2. Using the ratchet and the 9/16" socket, install the washer (D), lockwasher
(B), and bolt (A) to secure the upper frame to the hydraulic cylinder.
3. Using the torque wrench, torque the bolt (A) to 125 + 19 in-lb
(14.1 + 2.1 N·m).
4. If the stretcher is an OB/GYN Stretcher, replace the upper frame catch