4.28 OB/GYN Stretcher Foot Support Gas Spring
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 88
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
a. Using the ratchet and T30 Torx® head bit, remove the four screws (F)
securing the foot support bumper/handle (G) to the foot rest
weldment (E).
b. Remove the foot support bumper/handle (G) from the foot support
assembly.
c. Using the ratchet and T25 Torx® head bit, remove the two screws (H)
securing the mattress retainer (I) to the foot rest weldment (E).
d. Remove the mattress retainer (I) from the foot support assembly.
e. Using the side cutters, cut the ratchet fastener (J) off of the foot support
assembly.
5. Pull and remove the foot support bellow (A) back, and towards the foot end
of the stretcher to expose the foot support gas spring (B) (see figure 4-46
on page 4-88)
Figure 4-46. Foot Support Gas Spring
6. Remove and discard the remaining ratchet fastener from the release lever.
7. Using the adjustable wrench, loosen the jam nut (K) on the foot support gas
spring (B).
m115d202
A
Foot end
B
C
D
E
K
L
N
M
O