4.4 PC-100/PC-200 Procedural Stretcher Siderail Latch
Chapter 4: Removal, Replacement, and Adjustment Procedures
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
Page 4 - 17
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
4
4.4
PC-100/PC-200 Procedural Stretcher Siderail Latch
Tools required:
Ratchet
T30 Torx®
1
head bit
Torque wrench 0-250 in-lb
Removal
1. Remove the bottom rail end caps (A) from the bent bottom rail
extension (B).
2. Remove the side rail trim strip (C) from the bent bottom rail extension (B).
3. Using the ratchet and the T30 Torx® head bit, remove and discard the latch
shoulder bolt (D
)
that secures the latch (E) to the end tube (F) (see figure 4-
6 on page 4-17).
Figure 4-6. PC-100/PC-200 Procedural Stretcher Siderail Latch
4. Using the ratchet and the T30 Torx® head bit, remove the lower pivot bolt
(G), roller guide (H), latch bushing (I)
,
and latch (E) from the rail
assembly. Discard the lower pivot bolt and any damaged parts.
1. Torx® is a registered trademark of Textron, Inc.