4.6
PC-300–PC-550 Procedural/Fixed Height/Trauma/Surgical/
OB/GYN/Electric Stretcher Siderail Latch
Tools required:
Ratchet
T30 Torx®
1
head bit
Torque wrench 0-250 in-lb
Removal
1. Using the ratchet and the T30 Torx® head bit, remove the latch shoulder
bolt (D
)
that secures the latch plate (F) to the end tube (E) (see figure 4-8
on page 4-21).
Figure 4-8. Siderail Latch
2. Using the ratchet and the T30 Torx® head bit, remove the lower pivot bolt
(C), roller guide (B), latch bushing (A)
,
and latch plate (F) from the rail
assembly. Discard the lower pivot bolt and any damaged parts.
Replacement
1. Align the latch plate (F), the roller guide (B), and the latch bushing (A)
with the bolt holes in the rail assembly.
1. Torx® is a registered trademark of Textron, Inc.