4.6 PC-300–PC-550 Procedural/Fixed Height/Trauma/Surgical/ OB/GYN/Electric Stretcher Siderail
Latch
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 22
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
CAUTION:
The lower pivot bolt has self-tapping threads. Ensure that you use a
new lower pivot bolt for the installation procedure, and hand start it.
Failure to do so could result in damage to the bolt hole threads.
2. Using the ratchet and the T30 Torx® head bit, carefully install a
new
lower
pivot bolt (C)
to secure the latch plate (F) to the siderail.
3. Using the torque wrench, torque the lower pivot bolt (C)
to 140 + 10 in-lb
(15.8 + 1.1 N·m).
CAUTION:
The latch shoulder bolt has self-tapping threads. Ensure that you use a
new latch shoulder bolt for the installation procedure, and hand start it.
Failure to do so could result in damage to the bolt hole threads.
4. Using the ratchet and the T30 Torx® head bit, install a
new
latch shoulder
bolt (D)
to secure the latch plate (F) to the end tube (E).
5. Using the torque wrench, torque the latch shoulder bolt (D)
to
140 + 10 in-lb (15.8 + 1.1 N·m).