2.10 OB/GYN Stretcher Foot Supports Malfunction
Chapter 2: Troubleshooting Procedures
Page 2 - 50
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
3. The foot support rises without activating the release lever and does not lock
into position along its travel.
Yes
No
↓
→
Go to step 5.
4. Replace the defective foot support gas spring (refer to procedure 4.28). If
this solves the problem, go to “Final Actions” on page 2-32. Otherwise,
call Hill-Rom Technical Support at (800) 445-3720.
5. The foot support rotates outward and inward when the release handle at the
end of the foot support is activated.
Yes
No
↓
→
Replace the defective foot support Mechlok®
1
procedure 4.27). If this solves the problem, go to “Final
Actions” on page 2-32. Otherwise, call Hill-Rom Technical
Support at (800) 445-3720.
6. Go to “Final Actions” on page 2-32.