Electric Stretcher Function Checks
Chapter 2: Troubleshooting Procedures
Page 2 - 30
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
35. On the nurse control (see figure 2-2 on page 2-30) put the head ON/OFF
switch in the ON position. Push the head UP/DOWN switch to the DOWN
position. The head section lowers smoothly without binding.
Yes
No
↓
→
Figure 2-2. Nurse Control Switch
36. Release the head UP/DOWN switch when the desired position is reached.
The head section locks in place.
Yes
No
↓
→
Optional Push Handles Function
37. Raise the push handles. The push handles lock into position.
Yes
No
↓
→
38. Lift up on the push handle release latch. The push handles drop into the
stored position.
Yes
No
↓
→
m115a063