7.26 Integrated Fiber Optic Exam Light—P7915AT
Chapter 7: Accessories
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
Page 7 - 53
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
7
5. Using the adjustable wrench or 1½" open end wrench, remove the nut (L)
from the gooseneck assembly (H) (see figure 7-29 on page 7-53).
Figure 7-29. Gooseneck Assembly
6. Insert the fiber optic cable connector (M) through the available hole (N) at
the foot end of the stretcher.
7. Slide the nut (L) over the fiber optic cable connector (M), and hand tighten
the nut temporarily to secure the gooseneck assembly (H) to the stretcher.
8. Route the fiber optic cable assembly over the first crossmember and under
the catch shroud (O) toward the head end of the stretcher.
9. Using the tape measure, measure 1¼" (32 mm) from the top edge of the
welded crossmember (Q) toward the foot end of the stretcher.
10. Using a pencil, mark the correct measurement on the inside of the upper
frame fixed weldment (R).
L
H
m115d190
M
N
O
Head end
1¼"
(32 mm)
3½"
(89 mm)
Q
R
S
T