4.29 PC-100/PC-200 Procedural Stretcher Caster
Chapter 4: Removal, Replacement, and Adjustment Procedures
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
Page 4 - 91
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
4
4.29 PC-100/PC-200 Procedural Stretcher Caster
Tools required:
Ratchet
7/16" socket
T25 Torx®
1
head bit
Hammer
Phillips head screwdriver
5/16" pin punch
Torque wrench 0-250 in-lb
Blue Loctite®
2
(P/N SA3618) adhesive
8" x 2" x 4" (203 mm x 51 mm x 102 mm) lumber
Removal
1. Pull the base shroud free of its Velcro®
3
fastenings, and secure the base
shroud (A) to the upper frame (B) (see figure 4-47 on page 4-91).
Figure 4-47. PC-100/PC-200 Procedural Stretcher Shroud and Upper Frame
•
If patient security straps are available, use them to secure the base
shroud to the upper frame.
•
If the patient security straps are not available, you can use rope, or
you can remove the base shroud (refer to procedure 4.25).
1. Torx® is a registered trademark of Textron, Inc.
2. Loctite® is a registered trademark of Loctite Corporation.
3. Velcro® is a registered trademark of Velcro Industries, BV (a Dutch corporation).