7.26 Integrated Fiber Optic Exam Light—P7915AT
Chapter 7: Accessories
Page 7 - 50
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
7.26 Integrated Fiber Optic Exam Light—P7915AT
Tools required:
Ratchet
T25 Torx®
1
head bit
Tape measure
Box knife
Adjustable wrench or 1½" open end wrench
If you are installing the complete (non-factory installed) fiber optic exam light
assembly, go to “Complete (Non-Factory Installed) Installation” on page 7-52.
Otherwise, go to “Factory Installation” on page 7-50 to complete the factory
installation of the fiber optic exam light.
Factory Installation
1. Set the central brake pedal to the brake position.
2. Using the hilow operation pedals, raise the stretcher to the desired level
position.
3. Squeeze the back panel release handle, and raise the back section to
approximately 90°.
4. Remove the illuminator box (A) from its packaging container (see figure 7-
5. Orient the illuminator box (A) above the light mounting plate (B), so that
the power switch (C) is facing toward the foot end of the stretcher.
6. Lower the illuminator box (A) onto the light mounting plate (B), and insert
the light mounting plate tabs (D) into the keyhole slots (E) located on the
bottom of the illuminator box (A).
7. Slide the illuminator box (A) toward the head end of the stretcher to secure
the illuminator box (A) onto the light mounting plate (B).
8. Remove the power cord (X) from its packaging container.
9. Route the power cord (X) through the cable tie (F) on the light mounting
plate (B).
10. Plug the power cord (X) into the appropriate connection on the illuminator
box (A).
1. Torx® is a registered trademark of Textron, Inc.