2.8 Procedural/Fixed Height/Trauma/OB/GYN Stretcher Back Panel Does Not Raise/Lower Correctly
Chapter 2: Troubleshooting Procedures
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
Page 2 - 47
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
2
2.8
Procedural/Fixed Height/Trauma/OB/GYN Stretcher Back
Panel Does Not Raise/Lower Correctly
1. The headed pins (A), solid rivets (B), push nuts (C), and retaining rings (D)
are in place (see figure 2-10 on page 2-47).
Yes
No
↓
→
Replace any missing hardware. If this solves the problem, go to
“Final Actions” on page 2-32. Otherwise, continue to step 2.
Figure 2-10. Back Panel
2. When the release handle is activated, the back panel rises without
assistance.
Yes
No
↓
→
Replace the defective back panel gas spring (refer to
procedure 4.16). If this solves the problem, go to “Final
Actions” on page 2-32. Otherwise, call Hill-Rom Technical
Support at (800) 445-3720.
3. Go to “Final Actions” on page 2-32.