OB/GYN Stretcher Function Checks
Chapter 2: Troubleshooting Procedures
Page 2 - 20
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
position, the gas spring releases and locks throughout its full range of
motion, and the release handle engages and disengages completely,
returning to its original position with no evidence of binding. Do the same
thing for the
right
release handle.
NOTE:
The PACU extenders can either be removed from the stretcher, in the
armboard position, or in the extender position. If they are in the extender
position (around the articulating head section), then the release handle on the
PACU extenders must be used to raise or lower the back section.
Yes
No
↓
→
Headrest Function
28. Raise and lower the articulating headrest section, as well as rotate in both
directions (forward/backward, up/down) by engaging and disengaging the
thumb push lever located on either side of the headrest section. The
articulating headrest moves freely with no evidence of binding along both
axis. The gas spring releases and locks throughout its full range of motion,
and the push handle release engages and disengages completely, returning
to its original position with no evidence of binding.
Yes
No
↓
→
Thigh Section Function
29. Crank the knee Gatch. The knee Gatch rises and lowers evenly without
binding.
Yes
No
↓
→
30. Go to “Final Actions” on page 2-32.
OB/GYN Stretcher Function Checks
Brake/Steer Function
1. Initial Actions have been performed.
Yes