2.9 Procedural Stretcher Thigh Panel Will Not Raise/Lower
Chapter 2: Troubleshooting Procedures
Page 2 - 48
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
2.9
Procedural Stretcher Thigh Panel Will Not Raise/Lower
1. The crank handle (B) is attached to the knee screw assembly (A)
(see figure 2-11 on page 2-48).
Yes
No
↓
→
Attach the crank handle (refer to procedure 4.19). If this solves
the problem, go to “Final Actions” on page 2-32. Otherwise,
continue to step 2.
Figure 2-11. Knee Screw Assembly
2. Another part of the stretcher interferes with the knee screw assembly.
Yes
No
↓
→
Go to step 5.
3. Take appropriate actions to eliminate the interference.
This solves the problem.
Yes
No
↓
→
Go to step 5.
4. Go to “Final Actions” on page 2-32.
5. Replace the knee screw assembly (refer to procedure 4.19).
This solves the problem.
Yes
No
↓
→
Call Hill-Rom Technical Support at (800) 445-3720.
6. Go to “Final Actions” on page 2-32.
B
A
m115_013