7.7 IV Transporter
Chapter 7: Accessories
Page 7 - 14
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
7.7
IV Transporter
The IV transporter attaches to the base of the stretcher and allows for the
transport of a portable IV pole without the assistance of the caregiver.
The IV transporter can be mounted on either the right or left side of the
stretcher at the head end only.
Installation
Tools required:
Ratchet
¾" socket
Torque wrench 0-250 in-lb
Drill
¼" drill bit
7/8" drill bit
Retaining ring removal/installation tool
1. Decide on which side of the stretcher you will mount the transporter.
NOTE:
There are dimples on the inside surface on both sides of the head end of the
shroud. These identify the proper location for the transporter hole.
2. Lift the shroud, and locate the dimple at the head end of the shroud that
corresponds to the location where you want to mount the transporter.
3. Using the drill and the ¼" drill bit, drill a ¼" (6.4 mm) pilot hole through
the center of the dimple.
4. Using the drill with the 7/8" drill bit and using the pilot hole as a guide,
carefully drill a 7/8" (22.2 mm) hole through the shroud.
5. Install the rubber grommet (D) into the hole you just drilled in the shroud