2.11 OB/GYN Foot Section Does Not Release Correctly
Chapter 2: Troubleshooting Procedures
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
Page 2 - 51
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
2
2.11 OB/GYN Foot Section Does Not Release Correctly
1. Another part of the stretcher interferes with the foot section assembly.
Yes
No
↓
→
Go to step 4.
2. Take appropriate actions to eliminate the interference.
This solves the problem.
Yes
No
↓
→
Go to step 4.
3. Go to “Final Actions” on page 2-32.
4. The stretcher is in the standard position (non-examination).
Yes
No
↓
→
Slide the patient platform into the standard position. If this
solves the problem, go to “Final Actions” on page 2-32.
Otherwise, continue to step 5.
5. Check for proper operation of the release handle and cable function. All
mechanisms are working properly.
Yes
No
↓
→
Adjust the foot section release handle cable assembly. If this
solves the problem, go to “Final Actions” on page 2-32.
Otherwise, continue to step 6.
6. Replace the foot section assembly. The foot section operates properly.
Yes
No
↓
→
Call Hill-Rom Technical Support at (800) 445-3720.