2.5 Reduced Pedal Pumping
Chapter 2: Troubleshooting Procedures
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
Page 2 - 41
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
2
5. Lower the stretcher, and press the pump pedal 28 times (24 times for the
PC-100/PC-200 stretcher) to raise the stretcher to the high position. Both
the head and foot ends of the stretcher raise evenly.
Yes
No
↓
→
If the head end does not rise, replace the hydraulic cylinder at the
head end of the stretcher. If the foot end does not rise, replace the
hydraulic cylinder at the foot end of the stretcher (refer to
procedure 4.32). If this solves the problem, go to “Final
Actions” on page 2-32. Otherwise, call Hill-Rom Technical
Support at (800) 445-3720.