4.22 Electric Stretcher CPR Release Cable Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
Page 4 - 71
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
4
Adjustment
Using the ½" open end wrench, adjust the cable adjusting nut (I) (see figure 4-
4 on page 4-13) on the threaded stud such that the handle on the CPR handle
weldment just touches the plastic CPR handle when the CPR latch (L)
(see figure 4-33 on page 4-69) is fully extended.