4.30 PC-300–PC-550 Procedural/Fixed Height/Trauma/Surgical/ OB/GYN/Electric Stretcher Caster
Chapter 4: Removal, Replacement, and Adjustment Procedures
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
Page 4 - 95
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
4
4.30 PC-300–PC-550 Procedural/Fixed Height/Trauma/Surgical/
OB/GYN/Electric Stretcher Caster
Tools required:
Ratchet
Hammer
7/16" socket
5/16" pin punch
13 mm socket*
1/8" hex head bit
Torque wrench 0-250 in-lb
Blue Loctite®
1
(P/N SA3618) adhesive
8" x 2" x 4" (203 mm x 51 mm x 102 mm) lumber
*(surgical stretcher only)
Removal
1. Pull the base shroud (A) free of its Velcro®
2
fastenings, and secure it to the
upper frame (B) (see figure 4-50 on page 4-95).
Figure 4-50. Base Shroud Secured to Upper Frame
1. Loctite® is a registered trademark of Loctite Corporation.