4.27 OB/GYN Stretcher Foot Support Mechlok® Device
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 84
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
4.27 OB/GYN Stretcher Foot Support Mechlok®
1
Device
Tools required:
Ratchet
3/16" hex head bit
1/8" hex head bit
Screwdriver
Retaining ring removal/installation tool
Removal
1. Remove the mattress assembly from the stretcher.
WARNING:
Make sure the stretcher is in the full flat position, and make sure it is
secure. Failure to do so could result in personal injury or equipment
damage.
2. Put the stretcher in the full flat position.
3. Remove the foot section from the stretcher.
4. Using the screwdriver, remove the two screws (J) that secure the foot
section guide (K) to the upper frame carriage weldment (L) (see figure 4-
43 on page 4-84).
Figure 4-43. Foot Section Guide
1. Mechlok® is a registered trademark of P. L. Porter Company.
m115d205
J
K
L
Head end