4.27 OB/GYN Stretcher Foot Support Mechlok® Device
Chapter 4: Removal, Replacement, and Adjustment Procedures
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
Page 4 - 85
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
4
5. Slide the foot section guide (K) toward the head end of the stretcher, and
remove it from the stretcher.
6. Pull the foot support bellows (A) back and towards the foot end of the
stretcher to expose the foot support Mechlok® device (B) (see figure 4-44
on page 4-85).
Figure 4-44. Foot Support Mechlok® Device
7. Using the ratchet and 3/16" hex head bit, remove the shoulder bolt (C)
securing the foot support Mechlok® device (B) to the upper frame carriage
weldment (D).
8. Using the ratchet, 1/8" hex head bit and adjustable wrench, remove the
shoulder bolt (E) and locknut (F) securing the foot support Mechlok®
device (B) to the foot support base weldment (G).
9. Using the retaining ring removal/installation tool, remove the retaining ring
(H) that secures the foot support angle cable (I) to the foot support
Mechlok® device (B).
10. Remove the foot support Mechlok® device (B) from the stretcher.
m115d193
A
Foot end
B
C
D
E
F
G
H
I