2.14 Electric Stretcher Thigh Panel Does Not Raise/Lower Electrically
Chapter 2: Troubleshooting Procedures
Page 2 - 54
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
2.14 Electric Stretcher Thigh Panel Does Not Raise/Lower
Electrically
1. The stretcher has been connected to an external power supply, and the
power cord is not damaged.
Yes
No
↓
→
Go to “Initial Actions” on page 2-5.
2. On the nurse control, the head ON/OFF switch is in the ON position, and
the LEDs are lit (see figure 2-2 on page 2-30).
Yes
No
↓
→
Go to step 5.
3. Push the head up or down arrow on the patient control switch on both sides
(see figure 2-1 on page 2-28). The knee drive operates in both directions
for both sides.
Yes
No
↓
→
Check the previous steps. Go to step 5.
4. Go to “Final Actions” on page 2-32.
5. The plugs in the control board are all seated properly.
Yes
No
↓
→
Seat the plugs firmly in place. Check the previous steps. Go to
step 2. If this solves the problem, go to “Final Actions” on page
2-32. Otherwise, call Hill-Rom Technical Support at
(800) 445-3720.