Useful Function
------------------------------------------------------------------------------------
38
Variable Slew Rate
Sets the increasing and falling rate for each
output voltage and output current individually.
■Constant Voltage (CV) Slew Rate
Use when a small transition rate is desired in the
output voltage when using constant voltage.
Range:0.1V/s~160.0V/s
■Constant Current (CC) Slew Rate
Use when a small transition rate is desired in the
output voltage when using constant current.
Range:
400H/HA : 0.01A/s ~ 10.00A/s
800H/HA : 0.01A/s ~ 20.00A/s
1600H/HA : 0.01A/s ~ 40.00A/s
Setting Procedures
Press the Function “FUNC.” key.
FUNC. lamp will light.
Settings Item Number will display on the top “Number
Display” and parameters on the bottom Number Display.
Select Item Number 2 with the VOLTAGE dial.
Assign the Startup Priority Mode by setting the
parameter to 2 or 3 with the CURRENT dial.
Set Value 2=CV Priortiy, Variable Slew Rate
Set Value 3=CC Priority, Varialbe Slew Rate
The range will vary for the Constant Current(CC) Slew Rate depending
on the number of parallel devices.
Ex: For 400H/HA
① 2 Devices:0.01~20.00A/s
⑥ 7 Devices: 0.01~70.00A/s
② 3 Devices:0.01~30.00A/s
⑦ 8 Devices: 0.01~80.00A/s
③ 4 Devices:0.01~40.00A/s
⑧ 9 Devices: 0.01~90.00A/s
④ 5 Devices:0.01~50.00A/s
⑨ 10 Devices:0.01~100.0A/s
⑤ 6 Devices:0.01~60.00A/s
Caution
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
Set to 2 or 3
①CVスルーレート立ち上り
②CVスルーレート立ち下り
電圧(
V)
→
①
△V
△t
①
②
②
△V
△t
時間(s)→
CV優先モードでスルーレート設定をした例
スルーレート(CV)= (V/S)
△t
△V
Example of Slew Rate Settings in CV Priority Mode
Time (s)→
① CV Slew Rate Rising
② CV Slew Rate Falling
Slew Rate (CV)
Voltage
(V)
→
Summary of Contents for ZX-S-H Series
Page 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual ...
Page 37: ...Basic Operations 33 This is blank page ...
Page 59: ...Useful Function 55 This is blank page ...
Page 115: ...Using via Digital Data Communication 111 This is blank page ...
Page 141: ...Controlling with External Analog Contact Signals 137 This is blank page ...
Page 147: ...Special Loads 143 This is blank page 守 ...
Page 188: ... 184 付録 External Diagrams ZX S 400H 400HA 4 M3 ...
Page 189: ...Appendix 185 ZX S 800H 800HA ...
Page 190: ...Appendix 186 ZX S 1600H 1600HA ...