Calibration
------------------------------------------------------------------------------------
154
Output Voltage Control with External Resistance(A)
1. Disconnect the load and remote sensing wires and
connect the digital multimeter to the output terminal.
Turn the POWER switch ON and OUTPUT OFF in advance.
2. Press the FUNC. key and change the top Number
Display to Settings Item 70 with the VOLTAGE dial.
3. Change the set value to 2 with the CURRENT dial.
2: External Resistance(0~∞Ω A)
4. Press the FUNC. key again to finalize settings.
5. Press the OVP key ①. Set to “704.0V” with the
VOLTAGE dial ②.
6. Set external resistance to infinity (open).
7. Press OUTPUT key to set the output to ON.
Calibration Circuit
Digital
Multimeter
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
+
-
V
Power Supply
+
-
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
IN
OUT
18
6
外部抵抗
0~∞
Ω
External
Resistance
0~∞Ω
Set FUNCTION Settings Item 13, Linearity compensation
Function, to 0 (0=OFF) beforehand.
OUT PU T
PR ES ET
OVP
OCP
POWE R
FU NC.
LOCAL
LOCK
D ISPL AY
ST ORE
MEMOR Y
CC
CV
OUTPU T POWER
(%)
ST EP
CUR RE NT
ST EP
V OL TAGE
PRE SET FUN C. REMOTE LOCK
A LAR M R ESE T
20
60
100 LMT
40
80
①
②
ZX-S-4 00 L
A
B
C
Summary of Contents for ZX-S-H Series
Page 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual ...
Page 37: ...Basic Operations 33 This is blank page ...
Page 59: ...Useful Function 55 This is blank page ...
Page 115: ...Using via Digital Data Communication 111 This is blank page ...
Page 141: ...Controlling with External Analog Contact Signals 137 This is blank page ...
Page 147: ...Special Loads 143 This is blank page 守 ...
Page 188: ... 184 付録 External Diagrams ZX S 400H 400HA 4 M3 ...
Page 189: ...Appendix 185 ZX S 800H 800HA ...
Page 190: ...Appendix 186 ZX S 1600H 1600HA ...