Controlling with External Analog & Contact Signals
------------------------------------------------------------------------------------
136
LEVEL_ALM1 Settings
This handles the operation mode’s status signal as a LEVEL_ALM1 (alarm status).
It can also be used for surveillance as an alarm when switching from constant voltage to constant
current.
Status Signals that can be set in LEVEL1_ALM
CV_STS(Constant Voltage Operation)
CC_STS(Constant Current Operation)
PL_STS(Power Limiter Operation)
Setting Procedures
Press The Function “FUNC.” key.
FUNC. lamp will light.
Settings Item Number (top) and parameters (bottom) will appear
in the Number Display.
Select the Item number with the VOLTAGE dial.
Item 74 : CV_STS
Item 75 : CC_STS
Item 76 : PL_STS
Change the parameters with the CURRENT dial.
0=Not include in LEVEL_ALM
1=Included in LEVEL_ALM
Press the FUNC. key again to finalize settings.
Returns the Number Display to measurement display mode.
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
Summary of Contents for ZX-S-H Series
Page 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual ...
Page 37: ...Basic Operations 33 This is blank page ...
Page 59: ...Useful Function 55 This is blank page ...
Page 115: ...Using via Digital Data Communication 111 This is blank page ...
Page 141: ...Controlling with External Analog Contact Signals 137 This is blank page ...
Page 147: ...Special Loads 143 This is blank page 守 ...
Page 188: ... 184 付録 External Diagrams ZX S 400H 400HA 4 M3 ...
Page 189: ...Appendix 185 ZX S 800H 800HA ...
Page 190: ...Appendix 186 ZX S 1600H 1600HA ...