Calibration
------------------------------------------------------------------------------------
151
6.Press the FUNC. key to complete offset calibration 10.Adjust so that the digital multimeter (used for output
of the output current settings. current measurement) displays “25.00mV”(※)
with the CURRENT dial.
※For 400H/HA equivalent to 5A at 25.00mV
7.Press the PRESET key ① and set the CC set value
to below(※) with the CURRENT dial ②.
※For 400H/HA set to “5.000A”
For 800H/HA set to “10.00A”
For 1600H/HA set to “20.00A”
(See Operating as Constant Current Power Supply
under Basic Operations on how.)
8.Press the PRESET key.
9. Press the FUNC. key and change the top Number
Display to Settings Item 23 with the VOLTAGE dial.
For 800H/HA equivalent to 10A at 20.00mV
For 1600H/HA equivalent to 20A at 20.00mV
1 1 . Press the FUNC. key to complete full-scale
calibration of the output current settings.
1 2 . This completes offset/full-scale calibration of
Output Current Settings.
Please return the voltage/current set values back
to normal before use.
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
FUNC. lamp will light and the
display on the bottom Number
Display will light out.
MEMO
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
Be certain to perform this along with meter calibration
(Offset/Full-Scale Ammeter Calibration.)
Caution
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
OUTP UT
PR ESET
OVP
OCP
POWER
FU NC.
LOCA L
LOCK
DIS PLAY
A
B
C
ST OR E
MEMOR Y
CC
CV
OU TPUT PO WE R
(% )
ST EP
CURR ENT
ST EP
VOLTA GE
PRE SE T F UNC. RE M OTE L OC K
ALA RM R ESET
2 0
6 0
1 00 L MT
4 0
8 0
①
②
ZX-S-400 H
Summary of Contents for ZX-S-H Series
Page 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual ...
Page 37: ...Basic Operations 33 This is blank page ...
Page 59: ...Useful Function 55 This is blank page ...
Page 115: ...Using via Digital Data Communication 111 This is blank page ...
Page 141: ...Controlling with External Analog Contact Signals 137 This is blank page ...
Page 147: ...Special Loads 143 This is blank page 守 ...
Page 188: ... 184 付録 External Diagrams ZX S 400H 400HA 4 M3 ...
Page 189: ...Appendix 185 ZX S 800H 800HA ...
Page 190: ...Appendix 186 ZX S 1600H 1600HA ...