Basic Operations
------------------------------------------------------------------------------------
16
Operating as Constant Voltage Power Supply
Please verify that the settings for Over Voltage
Protection (OVP) circuit (page 29) and Over Current
Protection (OCP) circuit (page 31) are in effect.
Turn the POWER switch ON to startup the power supply.
The voltage measurement value will display in the top Number
Display, and the current measurement value will display in the
bottom Number Display. Lamp “V” and “A” will light. (If
parameters for FUNCTION settings item 52 is set to 1, PRESET
value will be displayed)
Press the PRESET key
PRESET lamp will light.
Press the PRESET key again if you wish to discontinue and
return to the measurement display mode.
The voltage PRESET value will display in the top Number Display
and the current limit PRESET value will display in the bottom
Number Display.
Turn the VOLTAGE dial to adjust the Constant Voltage
Set Value
(Turn the dial right to raise or left to lower the set value.)
If the parameter for FUNCTION item 50 is set to 0, press
the PRESET key again. The output voltage will be changed
once PRESET mode has been exited.
If the parameter for FUNCTION item 50 is set to 1, promptly
change the set value using the VOLTAGE dial.
Press the VOLTAGE dial to display the current number
position. The selected number will be brightened. Press to
change position. If all 4 places are brightened, numbers will
change from the lowest place.
MEMO
⇒Continued on next page
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
When POWER is turned OFF the top and bottom
Number Displays will be displayed in the same
manner when POWER is turned ON again.
MEMO
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
Summary of Contents for ZX-S-H Series
Page 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual ...
Page 37: ...Basic Operations 33 This is blank page ...
Page 59: ...Useful Function 55 This is blank page ...
Page 115: ...Using via Digital Data Communication 111 This is blank page ...
Page 141: ...Controlling with External Analog Contact Signals 137 This is blank page ...
Page 147: ...Special Loads 143 This is blank page 守 ...
Page 188: ... 184 付録 External Diagrams ZX S 400H 400HA 4 M3 ...
Page 189: ...Appendix 185 ZX S 800H 800HA ...
Page 190: ...Appendix 186 ZX S 1600H 1600HA ...