Calibration
------------------------------------------------------------------------------------
173
7.Press the FUNC. key and change the top Number
Display to Settings Item 32 with the VOLTAGE dial.
8.Adjust so that the digital multimeter No.2 (used for
voltage monitor measurement) displays “0.00V” with
the CURRENT dial.
9.Press the FUNC. key to complete offset calibration
of output voltage monitor.
10.Press the FUNC. key and change the top Number
Display to Settings Item 33 with the VOLTAGE dial.
11.Adjust so that the digital multimeter No. 2 (used for
voltage monitor measurement) displays “Output
Voltage Digital Multimeter No. 1/8” with the
CURRENT dial.
1 2 . Press the FUNC. key to complete full-scale
calibration of output voltage monitor.
13. This completes offset/full-scale calibration of
output voltage monitor.
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
Summary of Contents for ZX-S-H Series
Page 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual ...
Page 37: ...Basic Operations 33 This is blank page ...
Page 59: ...Useful Function 55 This is blank page ...
Page 115: ...Using via Digital Data Communication 111 This is blank page ...
Page 141: ...Controlling with External Analog Contact Signals 137 This is blank page ...
Page 147: ...Special Loads 143 This is blank page 守 ...
Page 188: ... 184 付録 External Diagrams ZX S 400H 400HA 4 M3 ...
Page 189: ...Appendix 185 ZX S 800H 800HA ...
Page 190: ...Appendix 186 ZX S 1600H 1600HA ...