Calibration
------------------------------------------------------------------------------------
174
Offset/Full-Scale Calibration of Output Current Monitor(ZX-S Series Type-HA)
1.Perform load with only the shunt resistor.
Disconnect the load wires and the remote sensing wires
.
Connect the shunt to the output terminal. Connect the
digital multimeter to the voltage terminal located on the
shunt.
Connect the digital multimeter to the external control
connectors number 3 and 4. (Current Monitor Terminal)
Turn the POWER switch ON and OUTPUT OFF in
advance.
2 . Press the FUNC. key and change the top Number
Display to Settings Item 13 with the VOLTAGE dial.
3.Turn the CURRENT dial and set the current value to 0
(Linearity Compensation: OFF).
4.Press the FUNC. key again to finalize settings.
5.Press the PRESET key ① and set to “1.00V” with
the VOLTAGE dial ② and “5.000A~20.0A” (※)
with the CURRENT dial ③.
※For 400H/HA set to “5.000A”
For 800H/HA set to “10.00A”
For 1600H/HA set to “20.0A”
6.Press the PRESET key ①
7. Set the load switch S1 to OPEN.
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
Note:
Current is 5A for
(400H/HA)
800H/HA is 10A
1600H/HA is 20A
Use wires which will not
produce a voltage drop
below 0.8V.
+
-
Digital
Multimeter No.1
Power
Supply
V
Shunt
-
+
Calibration Circuit
V
+
-
3
4
Digital
Multimeter No.2
OUT PU T
PR ES ET
OVP
OCP
POWE R
FU NC.
LOCAL
LOCK
DISPL AY
A
B
C
ST ORE
MEMOR Y
CC
CV
OUTPU T POWER
(%)
ST EP
CUR RE NT
ST EP
V OL TAGE
PRE SET FUN C. REMOTE LOCK
A LAR M R ESE T
20
60
100 LMT
40
80
①
②
③
ZX-S-400H
Summary of Contents for ZX-S-H Series
Page 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual ...
Page 37: ...Basic Operations 33 This is blank page ...
Page 59: ...Useful Function 55 This is blank page ...
Page 115: ...Using via Digital Data Communication 111 This is blank page ...
Page 141: ...Controlling with External Analog Contact Signals 137 This is blank page ...
Page 147: ...Special Loads 143 This is blank page 守 ...
Page 188: ... 184 付録 External Diagrams ZX S 400H 400HA 4 M3 ...
Page 189: ...Appendix 185 ZX S 800H 800HA ...
Page 190: ...Appendix 186 ZX S 1600H 1600HA ...