Using via Digital Data Communication
------------------------------------------------------------------------------------
70
OUTPut:DELay:OFF
:Sets Output OFF Delay Time
Function: Sets Output OFF Delay Time
Same as FUNCTION Settings Item 9
Format:OUTPut:DELay:OFF <NRf>
Type:SET command, QUERY command
Paramters: 0.00~99.99
Response: Same as parameters
Restrictions: Alarm response is returned for normal SET commands in the following conditions
・While operating as a series operation slave unit
・While operating as a parallel operation slave unit
・Sequence operation in progress
<Ex.>
OUTP:DEL:OFF 56.78
OUTP:DEL:OFF?
56.78
OUTPut:EXTernal:MODE
:
Sets operation mode for Output Contorl at External Contact Points
Function: Sets operation mode for Output Control at External Contact Points
Same as FUNCTION Settings Item 11
Format:OUTPut:EXTernal:MODE <NRf>
Type:SET command, QUERY command
Paramters: 0…Invalid
1…When close Output is ON, When open Output is OFF
2…When close normal operation, When open Output is OFF
Response: Same as parameters
Restrictions: Alarm response is returned for normal SET commands in the following conditions
・While operating as a parallel operation slave unit
・
Sequence operation in progress
<Ex.>
OUTP:EXT:MODE 1
OUTP:EXT:MODE?
1
Summary of Contents for ZX-S-H Series
Page 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual ...
Page 37: ...Basic Operations 33 This is blank page ...
Page 59: ...Useful Function 55 This is blank page ...
Page 115: ...Using via Digital Data Communication 111 This is blank page ...
Page 141: ...Controlling with External Analog Contact Signals 137 This is blank page ...
Page 147: ...Special Loads 143 This is blank page 守 ...
Page 188: ... 184 付録 External Diagrams ZX S 400H 400HA 4 M3 ...
Page 189: ...Appendix 185 ZX S 800H 800HA ...
Page 190: ...Appendix 186 ZX S 1600H 1600HA ...