Controlling with External Analog & Contact Signals
------------------------------------------------------------------------------------
134
Analog Output Monitor (ZX-S Series Type-HA)
This can produce the DC voltage, which is proportional with the output voltage or output current.
Use when wishing to surveillance the output with an external meter or recorder.
Outputs a “0~10V” DC for output voltage/currents of “0 to rated output”.
Please use an input impedance of a 10kΩ or more for the external meter.
Output Voltage Monitor
The accuracy for the output voltage monitor is 0.2%±2mV
Connecting
Connect the meter or recorder to accessory connector terminals No. 4 and 15.
Output Current Monitor
The accuracy for the output current monitor is 0.5%±2mV
Connecting
Connect the meter or recorder to accessory connector terminals No. 3 and 4.
IN
OUT
4
15
+
-
V
IN
OUT
4
3
+
-
V
Output current wave forms cannot be observed by the current monitor.
Control Common (No. 4 pin) is internally connected to the minus side of the output terminal.
Caution
Control Common (No. 4 pin) is internally connected to the minus side of the output terminal.
Caution
Summary of Contents for ZX-S-H Series
Page 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual ...
Page 37: ...Basic Operations 33 This is blank page ...
Page 59: ...Useful Function 55 This is blank page ...
Page 115: ...Using via Digital Data Communication 111 This is blank page ...
Page 141: ...Controlling with External Analog Contact Signals 137 This is blank page ...
Page 147: ...Special Loads 143 This is blank page 守 ...
Page 188: ... 184 付録 External Diagrams ZX S 400H 400HA 4 M3 ...
Page 189: ...Appendix 185 ZX S 800H 800HA ...
Page 190: ...Appendix 186 ZX S 1600H 1600HA ...