Using via Digital Data Communication
------------------------------------------------------------------------------------
83
[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]
:Sets Output Voltage
Function:Sets the output voltage
Format:[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude] <NRf>
Type:SET command, QUERY command
Parameters: 0.00~672.0
Response:Same as parameters
Restrictions: Alarm response is returned for normal SET commands in the following conditions
・While operating as a series operation slave unit
・While operating as a parallel operation slave unit
・
Sequence operation in progress
<Ex.>
VOLT 640
VOLT?
640.0
[SOURce]:VOLTage:PROTection[:LEVel]
:Sets OVP Voltage
Function:Sets the OVP voltage
Format:[SOURce]:VOLTage:PROTection[:LEVel] <NRf>
Type:SET command, QUERY command
Parameters: 10~704
Response:Same as parameters
Restrictions: Alarm response is returned for normal SET commands in the following conditions
・While operating as a parallel operation slave unit
・
Sequence operation in progress
<Ex.>
VOLT:PROT 704
VOLT:PROT?
704.0