Connecting Loads
For wiring to the output terminal, please attach the M5
crimp-style terminal and be sure to tightly fasten the M5
screws.(※)
・Twisting wires can minimize the ripple and noise at
the load terminal.
・The noise level can be minimized smaller than the
standard value by connecting the C1 and C2 close to
the load terminal.
Using a small, high-frequency impedance for C1 and
C2, cut the lead wire as short as possible and
connect.
C1:Electrolytic Capacitor 100~1000μF
C2:Film Capacitor 1~10μF
MEMO
Please attach the terminal cover after wiring.
Mandatory
M5 Bolt
(includes spring and flat washers)
When grounding, use a load terminal
!
負荷
より合わせる
M5圧着端子
配線材
+
-
C1
C2
Load
Twisted
Wires
M5 Crimp-Style Terminal
Terminal Cover