![Rohde & Schwarz FSW13 User Manual Download Page 778](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_1477847778.webp)
Remote Commands
R&S
®
FSW
778
User Manual 1173.9411.02 ─ 19
CALC:PSE:SUBR 10
//Sets 10 peaks per range to be stored in the list.
//--------------Performing the Measurement-----
INIT:SPUR; *WAI
//Performs a spurious emission measurement and waits until the sweep has finished.
//---------------Retrieving Results-------------
CALC:LIM1:FAIL?
//Queries the result of the check for limit line 1.
TRAC? SPUR
//Queries the peak list of the spurious emission measurement.
11.5.8 Analyzing Statistics (APD, CCDF)
All remote control commands specific to statistical measurements are described here.
Activating Statistical Measurements
..................................................................... 778
Configuring Statistical Measurements
...................................................................779
Using Gate Ranges for Statistical Measurements
................................................ 780
..............................................................................................782
Performing a Statistical Measurement
.................................................................. 784
.................................................................................................784
Programming Example: Measuring Statistics
....................................................... 786
11.5.8.1
Activating Statistical Measurements
The following commands activate statistical measurements.
CALCulate<n>:STATistics:APD[:STATe]
.......................................................................... 778
CALCulate<n>:STATistics:CCDF[:STATe]
........................................................................ 778
CALCulate<n>:STATistics:APD[:STATe]
<State>
This command turns the APD measurement on and off.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:STAT:APD ON
Switches on the APD measurement.
Manual operation:
See
CALCulate<n>:STATistics:CCDF[:STATe]
<State>
This command turns the CCDF on and off.