![Rohde & Schwarz FSW13 User Manual Download Page 983](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_1477847983.webp)
Remote Commands
R&S
®
FSW
983
User Manual 1173.9411.02 ─ 19
Manual operation:
See
CALCulate:LIMit<k>:UPPer:MARGin
<Margin>
This command defines an area around an upper limit line where limit check violations
are still tolerated.
Parameters:
<Margin>
numeric value
*RST:
0
Default unit: dB
Manual operation:
See
CALCulate:LIMit<k>:UPPer:MODE
<Mode>
This command selects the vertical limit line scaling.
Parameters:
<Mode>
ABSolute
Limit line is defined by absolute physical values.
The unit is variable.
RELative
Limit line is defined by relative values related to the reference
level (dB).
*RST:
ABSolute
Manual operation:
See
CALCulate:LIMit<k>:UPPer:OFFSet
<Offset>
This command defines an offset for a complete upper limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Parameters:
<Offset>
Numeric value.
*RST:
0
Default unit: dB
Manual operation:
See
CALCulate:LIMit<k>:UPPer:SHIFt
<Distance>
This command moves a complete upper limit line vertically.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Analyzing Measurements (Basics)