Remote Commands
R&S
®
FSW
895
User Manual 1173.9411.02 ─ 19
<SerialNo>
Serial number of the power sensor assigned to the specified
index
Example:
SYST:COMM:RDEV:PMET2:DEF '','NRP-Z81','',
'123456'
Assigns the power sensor with the serial number '123456' to the
configuration "Power Sensor 2".
SYST:COMM:RDEV:PMET2:DEF?
Queries the sensor assigned to "Power Sensor 2".
Result:
'','NRP-Z81','USB','123456'
The NRP-Z81 power sensor with the serial number '123456' is
assigned to the "Power Sensor 2".
Manual operation:
See
Configuring Power Sensor Measurements
CALibration:PMETer<p>:ZERO:AUTO ONCE
...................................................................895
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]
........................................................... 896
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]:AUTO ONCE
....................................... 896
CALCulate<n>:PMETer<p>:RELative:STATe
....................................................................896
[SENSe:]PMETer<p>:DCYCle[:STATe]
............................................................................ 897
[SENSe:]PMETer<p>:DCYCle:VALue
.............................................................................. 897
...................................................................................898
[SENSe:]PMETer<p>:FREQuency:LINK
...........................................................................898
.......................................................................................... 899
[SENSe:]PMETer<p>:MTIMe:AVERage:COUNt
................................................................ 899
[SENSe:]PMETer<p>:MTIMe:AVERage[:STATe]
...............................................................899
[SENSe:]PMETer<p>:ROFFset[:STATe]
...........................................................................900
.........................................................................................900
[SENSe:]PMETer<p>:UPDate[:STATe]
............................................................................ 901
......................................................................................... 901
............................................................................... 901
CALibration:PMETer<p>:ZERO:AUTO ONCE
This commands starts to zero the power sensor.
Note that you have to disconnect the signals from the power sensor input before you
start to zero the power sensor. Otherwise, results are invalid.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
ONCE
Setting Basic Measurement Parameters