![Rohde & Schwarz FSW13 User Manual Download Page 219](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_1477847219.webp)
Measurements
R&S
®
FSW
219
User Manual 1173.9411.02 ─ 19
PMin/PMax
Defines the power limits for each power class. The first range always starts at -200
dBm (-INF) and the last range always stops at 200 dBm (+INF). These fields cannot be
modified. If more than one power class is defined, the value of "PMin" must be equal to
the value of "PMax" of the previous power class and vice versa.
Note that the power level may be equal to the lower limit(s), but must be lower than the
upper limit(s):
P
min
≦
P
<
P
max
Otherwise the ranges are corrected automatically.
Remote command:
Sweep List
Switches to the "Sweep List" tab of the "Spectrum Emission Mask" dialog box and
focuses the "Limit Check" setting for the corresponding power class (1-4) in the refer-
ence range (see
Adding or Removing a Power Class
Adds a new power class at the end of the list or removes the last power class. After
adding or removing, the last power class is adapted to end at "+INF". Note that a maxi-
mum of 4 power classes are available.
Remote command:
5.6.5.5
MSR Settings
In the "MSR Settings" tab of the "Spectrum Emission Mask" dialog box you configure
multi-standard radio (MSR) measurements, which allow you to perform SEM tests on
multiple carriers using different digital standards.
For details see
chapter 5.6.4.4, "Multi-Standard Radio (MSR) SEM Measurements"
Spectrum Emission Mask (SEM) Measurement