![Rohde & Schwarz FSW13 User Manual Download Page 1050](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_14778471050.webp)
Remote Commands
R&S
®
FSW
1050
User Manual 1173.9411.02 ─ 19
......................................................................................... 1052
..........................................................................................1052
SYSTem:HPCoupling
<CouplingType>
Controls the default coupling ratios in the HP emulation mode for:
●
span and resolution bandwidth (Span/RBW) and
●
resolution bandwidth and video bandwidth (RBW/VBW)
For FSP (=FSW), the standard parameter coupling of the instrument is used. As a
result, in most cases a shorter sweep time is used than in case of HP.
This command is only available if a HP language is selected using
Parameters:
<CouplingType>
HP | FSP
*RST:
FSP
Example:
SYSTem:HPC HP
Manual operation:
See
SYSTem:IFGain:MODE
<Mode>
Configures the internal IF gain settings in HP emulation mode due to the application
needs. This setting is only taken into account for resolution bandwidth < 300 kHz and is
only available if a HP language is selected using
.
Parameters:
<Mode>
NORM | PULS
NORM
Optimized for high dynamic range, overload limit is close to ref-
erence level.
PULS
Optimized for pulsed signals, overload limit up to 10 dB above
reference level.
*RST:
NORM
Example:
SYST:IFG:MODE PULS
Manual operation:
See
SYSTem:LANGuage
<Language>
This command defines the system language.
For details see
chapter 10.2, "GPIB Languages"
Note
: as of firmware version 1.60, this command is also used to emulate previous R&S
signal and spectrum analyzers, making the
SYST:COMP
command obsolete.
Emulating Other Instruments' Commands