![Rohde & Schwarz FSW13 User Manual Download Page 238](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_1477847238.webp)
Measurements
R&S
®
FSW
238
User Manual 1173.9411.02 ─ 19
File contents
Explanation
x-Unit;Hz;
y-Unit;dBm;
List evaluation settings
Margin;200;
Peak List margin
Reference range settings
RefType; CPOWER;
Reference power type
TxBandwidth;3840000;;Hz
Channel power settings
Filter State; ON;
Alpha;0.22;
PeaksPerRange;1;
Max. number of peaks per range to be detected
Values;2;
Number of detected peaks
File data section
0;-12750000;-2515000;30000;13242367500;-43.844
722747802734;-0.33028793334960938;49.6697120
66650391;FAIL;
Measured peak values:
<range number>;
<start frequency>;
<stop frequency>;
<resolution bandwidth of range>;
<frequency of peak>;
<absolute power in dBm of peak>;
<relative power in dBc of peak>; (related to the
channel power)
<distance to the limit line in dB>; (positive value
means above the limit)
<limit fail (pass = 0, fail =1)>;
2;2515000;12750000;30000;13257632500;-43.8447
22747802734;-0.33028793334960938;49.66971206
6650391;FAIL;
5.7 Spurious Emissions Measurement
The R&S
FSW supports Spurious Emissions measurements.
........................................................................................238
Spurious Emissions Measurement Results
...........................................................239
...................................................................................240
Spurious Emissions Measurement Configuration
................................................. 242
How to Perform a Spurious Emissions Measurement
...........................................248
Reference: ASCII Export File Format (Spurious)
..................................................249
5.7.1 About the Measurement
The Spurious Emissions measurement monitors unwanted RF products outside the
assigned frequency band generated by an amplifier. The spurious emissions are usu-
Spurious Emissions Measurement