Common Analysis and Display Functions
R&S
®
FSW
523
User Manual 1173.9411.02 ─ 19
Fig. 7-15: Fundamental wave and the frequency and level reference point
6. Make the step size for the center frequency correspond to the signal frequency: in
the "Frequency" configuration dialog box, select "Center Frequency Stepsize =
Marker".
The step size for the center frequency is now equal to the marker frequency.
7. Move the center frequency to the 2
nd
harmonic of the signal by pressing the UP
( ) key on the front panel.
The center frequency is set to the 2
nd
harmonic.
8. Place the delta marker on the 2
nd
harmonic: in the "Marker To" menu, select the
"Peak" softkey.
The delta marker moves to the maximum of the 2
nd
harmonic. The displayed level
result is relative to the reference point level (= fundamental wave level).
The other harmonics are measured by repeating steps
and
, with the cen-
ter frequency being incremented or decremented in steps of 128 MHz using the UP or
DOWN keys.