![Rohde & Schwarz FSW13 User Manual Download Page 714](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_1477847714.webp)
Remote Commands
R&S
®
FSW
714
User Manual 1173.9411.02 ─ 19
If you have defined an absolute limit as well as a relative limit, the R&S
FSW uses the
lower value for the limit check.
Suffix:
<gap>
.
1 | 2
Gap (CACLR) channel number
Parameters:
<Limit>
Defines the absolute limit of the specified gap channel in dBm.
<Reserved>
Ignored.
Example:
CALC:LIM:ACP:GAP2:ABS 44.2dBm,0
Manual operation:
See
CALCulate:LIMit:ACPower:GAP<gap>:ABSolute:STATe
<State>
This command turns the absolute limit check for the specified gap (CACLR) channel on
and off.
You have to activate the general ACLR limit check before using this command with
.
Suffix:
<gap>
.
1 | 2
Gap (CACLR) channel number
Parameters:
<State>
ON | OFF
*RST:
OFF
Manual operation:
See
CALCulate:LIMit:ACPower:GAP<gap>[:RELative]
<Limit>, <Reserved>
This command defines the relative limit of the specified gap (CACLR) channel. The ref-
erence value for the relative limit is the measured channel power.
If you have defined an absolute limit as well as a relative limit, the R&S
FSW uses the
lower value for the limit check.
Suffix:
<gap>
.
1 | 2
Gap (CACLR) channel number
Parameters:
<Limit>
Defines the relative limit of the specified gap channel in dB.
<Reserved>
Ignored.
Example:
CALC:LIM:ACP:GAP1:REL 3dB,0
Manual operation:
See
Configuring and Performing Measurements