![Rohde & Schwarz FSW13 User Manual Download Page 25](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_1477847025.webp)
Getting Started
R&S
®
FSW
25
User Manual 1173.9411.02 ─ 19
For details on the connector refer to
chapter 3.2.2.2, "AC Power Supply Connection
► Connect the R&S
FSW to the AC power supply using the supplied power cable.
Since the instrument is assembled in line with the specifications for safety class
EN61010, it may only be connected to an outlet that has a ground contact.
3.1.1.5
Switching the Instrument On and Off
Switching the instrument on
► Press the AC power switch on the rear panel to position "I".
The instrument is supplied with AC power. After booting, the instrument is ready for
operation. A green LED above the POWER key indicates this.
An orange LED indicates the instrument is in standby mode
Warm-up time for OCXO
When the instrument is switched on, the OCXO requires an extended warm-up time
(see data sheet). No warm-up time is required when starting from standby mode.
Switching the instrument off
1. Press the POWER key on the front panel.
The R&S
FSW switches to standby mode.
2. Set the AC power switch on the rear panel to position "O", or disconnect the instru-
ment from the AC power supply.
The R&S
FSW changes into off mode.
Risk of losing data
If you switch off the running instrument using the rear panel switch or by disconnecting
the power cord, the instrument loses its current settings. Furthermore, program data
may be lost.
Press the Power key first to shut down the application properly.