[SENSe:]ESPectrum<sb>:RANGe<range>:LIMit<PClass>:RELative:STARt:
FUNCtion
<Function>
This command enables the use of a function when defining the relative limit for a SEM
range.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<range>
1..30
Selects the SEM range.
<PClass>
1..4
Power class for which the limit is defined.
Parameters:
<Function>
Defines the function to be used to determine the relative limit
line start value
MAX
The maximum of the relative
and
the absolute level is used as
the limit start value. Use the
these values.
OFF
No function is used, the relative limit line is defined by a fixed
relative start value. Use the
mand to define this value.
*RST:
OFF
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STARt:
FUNCtion MAX
For a detailed example see
chapter 11.5.6.11, "Example: SEM
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<range>:LIMit<PClass>:RELative:STOP
<Limit>
This command defines a relative limit for a SEM range.
Unlike manual operation, you can define a relative limit anytime and regardless of the
limit check mode.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<range>
1..30
Selects the SEM range.
Configuring and Performing Measurements