CALCulate:LIMit:ACPower:GAP<gap>[:RELative:]STATe
<State>
This command turns the relative limit check for the specified gap (CACLR) channel on
and off.
You have to activate the general ACLR limit check before using this command with
.
Suffix:
<gap>
.
1 | 2
Gap (CACLR) channel number
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:LIM:ACP:GAP1:REL:STAT ON
Manual operation:
See
CALCulate:LIMit:ACPower:GAP<gap>:RESult?
This command queries the state of the limit check for
all
gap (CACLR) channels in an
MSR ACLR measurement.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps. See also
The results of the power limit checks are also indicated in the
STAT:QUES:ACPL
sta-
tus registry (see
"STATus:QUEStionable:ACPLimit Register"
Suffix:
<gap>
.
1 | 2
irrelevant
Return values:
<LowerGap1>,
<UpperGap1>,
<LowerGap2>,
<UpperGap2>
The command returns the results for the four gap channels for
each gap (max. 4).
PASSED
Limit check has passed.
FAIL
Limit check has failed.
NONE
No results available, e.g. because limit checking was deactiva-
ted
Example:
INIT:IMM;*WAI;
CALC:LIM:ACP:GAP2:RES?
PASSED,PASSED,
Usage:
Query only
Manual operation:
See
Configuring and Performing Measurements