Applications and Operating Modes
R&S
®
FSW
106
User Manual 1173.9411.02 ─ 19
ment, the sweep count is considered. Thus, each measurement may consist of several
sweeps. The currently active measurement is indicated by a symbol in the tab label.
The result displays of the individual channels are updated in the tabs as the measure-
ments are performed. Sequential operation itself is independent of the currently
dis-
played
tab.
Sequencer modes
Three different Sequencer modes are available:
●
Single Sequence
Similar to single sweep mode; each measurement is performed once, until all
measurements in all active channels have been performed.
●
Continuous Sequence
Similar to continuous sweep mode; the measurements in each active channel are
performed one after the other, repeatedly, in the same order, until sequential oper-
ation is stopped. This is the default Sequencer mode.
●
Channel-defined Sequence
First, a single sequence is performed. Then, only channels in continuous sweep
mode are repeated continuously.
Example: Sequencer procedure
Assume the following active channel definition:
Tab name
application
Sweep mode
Sweep count
Spectrum
Spectrum
Cont. Sweep
5
Spectrum 2
Spectrum
Single Sweep
6
Running a Sequence of Measurements