![Rohde & Schwarz FSW13 User Manual Download Page 855](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_1477847855.webp)
Remote Commands
R&S
®
FSW
855
User Manual 1173.9411.02 ─ 19
Parameters:
<Source>
IMMediate
Free Run
EXTernal
Trigger signal from the TRIGGER INPUT connector.
EXT2
Trigger signal from the TRIGGER INPUT/OUTPUT connector.
Note: Connector must be configured for "Input".
EXT3
Trigger signal from the TRIGGER 3 INPUT/ OUTPUT connector.
Note: Connector must be configured for "Input".
RFPower
First intermediate frequency
IFPower
Second intermediate frequency
TIME
Time interval
VIDeo
Video mode is available in the time domain and only in the
Spectrum application.
PSEN
External power sensor
*RST:
IMMediate
Example:
TRIG:SOUR EXT
Selects the external trigger input as source of the trigger signal
Manual operation:
See
"Using the power sensor as an external trigger"
on page 327
See
See
See
See
See
See
See
See
See
TRIGger[:SEQuence]:TIME:RINTerval
<Interval>
This command defines the repetition interval for the time trigger.
Parameters:
<Interval>
2.0 ms to 5000
Range:
2 ms to 5000 s
*RST:
1.0 s
Setting Basic Measurement Parameters