Common Measurement Settings
R&S
®
FSW
396
User Manual 1173.9411.02 ─ 19
.......................................................................... 397
Setting the Reference Level Automatically (Auto Level)
...............................397
..............................................................................398
Using Electronic Attenuation (Option B25)
..................................................................398
.............................................................................. 399
Reference Level
Defines the expected maximum reference level. Signal levels above this value may not
be measured correctly, which is indicated by the "IF OVLD" status display ("OVLD" for
analog baseband or digitial baseband input).
Defines the expected maximum reference level. Signal levels above this value may not
be measured correctly, which is indicated by the "IF OVLD" status display.
The reference level is also used to scale power diagrams; the reference level is then
used as the maximum on the y-axis.
Since the R&S
FSW hardware is adapted according to this value, it is recommended
that you set the reference level close above the expected maximum signal level to
ensure an optimum measurement (no compression, good signal-to-noise ratio).
Note that the "Reference Level" value ignores the
. It is
important to know the actual power level the R&S
FSW must handle.
For details see
chapter 6.4.1.1, "Reference Level"
Note that for input from the External Mixer (R&S
FSW-B21) the maximum reference
level also depends on the conversion loss, see
Remote command:
Amplitude and Vertical Axis Configuration