Remote Commands
R&S
®
FSW
845
User Manual 1173.9411.02 ─ 19
Parameters:
<Unit>
DBM | V | A | W | DBPW | WATT | DBUV | DBMV | VOLT |
DBUA | AMPere
*RST:
dBm
Example:
CALC:UNIT:POW DBM
Sets the power unit to dBm.
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:RLEVel
<ReferenceLevel>
This command defines the reference level.
With a reference level offset
≠
0, the value range of the reference level is modified by
the offset.
Parameters:
<ReferenceLevel>
The unit is variable.
Range:
see datasheet
*RST:
0 dBm
Example:
DISP:TRAC:Y:RLEV -60dBm
Usage:
SCPI confirmed
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:RLEVel:OFFSet
<Offset>
This command defines a reference level offset.
Parameters:
<Offset>
Range:
-200 dB to 200 dB
*RST:
0dB
Example:
DISP:TRAC:Y:RLEV:OFFS -10dB
Manual operation:
See
"Shifting the Display (Offset)"
See
"Shifting the Display (Offset)"
[SENSe:]POWer:NCORrection
<State>
This command turns noise cancellation on and off.
If noise cancellation is on, the R&S
FSW performs a reference measurement to deter-
mine its inherent noise and subtracts the result from the channel power measurement
result (first active trace only).
For more information see
Parameters:
<State>
ON | OFF
*RST:
OFF
Setting Basic Measurement Parameters